Stability indicating RP-HPLC method development and validation for the simultaneous estimation of Grazoprevir and Elbasvir in bulk and pharmaceutical dosage form
Keywords:
Grazoprevir, Elbasvir, RP-HPLCAbstract
A simple, Accurate and precise method was developed for the simultaneous estimation of the Grazoprevir and Elbasvir in Tablet dosage form. Chromatogram was run through Kromosil C18 (250 x 4.6 mm), 5m. Mobile phase containing Buffer: Acetonitrile taken in the ratio 45:55 was pumped through column at a flow rate of 1 ml/min. Buffer used in this method was Di Potassium Hydrogen ortho Phosphate. Temperature was maintained at 30°C. Optimized wavelength selected was 215 nm. Retention time of Elbasvir and Grazoprevir and were found to be 2.503 min and 3.004. %RSD of the Elbasvir and Grazoprevir were and found to be 0.3 and 0.4 respectively. %Recovery was obtained as 98.17% and 99.83% for Grazoprevir and Elbasvir respectively. LOD, LOQ values obtained from regression equations of Grazoprevir and Elbasvir were 0.24, 0.73 and 0.06, 0.19 respectively. Retention times were decreased and run time was decreased, so the method developed was simple and economical that can be adopted in regular Quality control test in Industries.